Philips-P-817-Service-Manual电路原理图.pdf
《Philips-P-817-Service-Manual电路原理图.pdf》由会员分享,可在线阅读,更多相关《Philips-P-817-Service-Manual电路原理图.pdf(11页珍藏版)》请在收音机爱好者资料库上搜索。
1、 PHILIPS DIRECTIONS FOR USE UNIVERSAL MEASURING INSTRUMENT P 817 00/01 66 580 33-3-120 1/361 2 CONTENTS APPLICATIONS 3 ELECTRICAL DATA 3 Protection 3 TRANSPORT 5 ACCESSORIES 5 MAINTENANCE 5 OPERATION 5 Zero Setting 5 Measuring 6 A. Direct voltage 6 B. Direct Current 6 C. Alternating voltage 7 D. Alt
2、ernating Current 7 E. Resistance 7 Db-Scale 7 REPLACEMENT OF BATTERIES 8 SOME EXAMPLES OF MEASUREMENTS 8 A. Direct-voltage measurements 8 B. D.C. Measurements 8 C. A.C. Measurements 9 METER CIRCUITS FOR THE VARIOUS MEASURING RANGES 9 CIRCUIT DIAGRAMS 10 RECOMMENDED BATTERIES 11 3 APPLICATIONS Betwee
3、n 50 and 70%of all electrical faults occurring in a radio receiver or amplifier are such that they can he easily traced by measurements, for which particularly the PHILIPS Universal Measuring Instrument P 817 00/01 is eminently suitable. It is the ideal measuring instrument for use in the smaller re
4、pair shops or at the customers home, owing to its small dimensions and to the fact it does not require any mains supply. The following measurements can be carried out with the meter: a. direct and alternating voltages from 0 to 1200 V; b. direct and alternating currents from 0 to 3 A; c. resistances
5、 from 0 to 10 M ELECTRICAL DATA The voltage loss and the current consumption as given in the table at page 4 apply to full deflection; at less deflection these values are proportionally lower. All current and voltage scales are linear. For alternating-current and alternating-voltage measurements the
6、 reading remains accurate throughout the frequency range 30 10,000 c/s, with the exception of the measuring range of 1200 V, for which the upper accuracy limit lies at 5,000 c/s. The A.C. and alternating-voltage ranges have been calibrated for a purely sinusoidal signal. Although the meter can, if n
7、ecessary, be used standing upright, it measures more accurately lying down (see also page 5, under “zero setting“). PROTECTION For all commonly used measuring ranges the galvanometer can stand overloading for a short time. This does not hold, however, if the knob Sk2 is set to the very sensitive pos
8、ition “60mV“; in this position the protection is cut out and the meter cannot stand overload any longer. The maximum current is 30 A. 4 Measurement Measuring range Position Of Sk1 Position Of Sk2 Scale Current consumption or voltage loss at full deflection Sensitivity V 0 60 mV Any position 60 mV *
9、AV 33,300 /V 0 3 V 0 12 V 0 30 V 0 120 V 0 300 V 0 -1200 V 3 V 12 V 30 V 120 V 300 V 1200 V 30 A 25 A 40,000 /V I 0 30 A Any position 60 mV * AV 60 mV 0 120 A 0 600 A 0 6 mA 0 60 mA 0 600 Ma 0 3 A 0.12 mA 0.6 mA 6 mA 60 mA 600 mA 3 A 285 mV 345 mV 360 mV 360 mV 400 mV 580 mV V 0 3V 0 12 V 0 30 V 0 1
10、20 V 0 300 V 0 1200 V 3 V 12 V 30 V 120 V 300 V 1200 V AV 5 Ma 600 A 1666 /V I 0 - 600 A 0 6 mA 0 60 mA 0 600 mA 0 3 A 0.6 mA 6 mA 60 mA 600 mA 3 A AV R 0 1000 0 100 k 0 10 M Any position X 1 X 100 X 10 k 83 mA 0.8 mA 0.a3 mA Centre-scale Value 18 1800 180,000 5 TRANSPORT The right-hand knob (Sk2) i
11、s provided with a green dot. Whenever the instrument is transported, set this knob to the position marked with the green dot. In this position the galvanometer is short-circuited. The same precaution should he taken when the meter is to be sent to the repair shop. ACCESSORIES (available on order) 1
12、pair of universal plugs, type P 995 00; 1 pair of measuring flexes, type P 997 04, complete with probe pins; 1 leather bag with carrying straps, type P 997 05; 1 transistor tester, type P 997 09; 1 step down transformer (300/3 Amps, 2 VA, class 1, type P 997 10. MAINTENANCE The inside of the window
13、in the front panel of the universal meter is coated with a colourless liquid having antistatic properties, to prevent static charges, which may affect the position of the pointer and attract dust. This side of the window must not be rubbed with a cloth or anything similar, since otherwise the inside
14、 will become dull and the antistatic effect will be lost. OPERATION The position of knobs and sockets is shown in Fig. 1. ZERO SETTING Before starting measurements, it is recommended to check and, if necessary, to correct the zero settings as follows: a. for the current/voltage scale (A V), mechanic
15、ally with the set screw “St“; b. for the resistance scale (), with knob “R“, with sockets + and - short-circuited (see also point E at page 7). Remark concerning the zero adjustment Turning over a meter that has been checked for use in the flat position to the upright position (or the reverse) may r
16、esult in a difference of one scale division in the deflection of the pointer at any point on the 0 - 12 V scale; it is therefore necessary always to check the zero adjustment again and, if necessary, to correct it. 6 MEASURING A. Direct voltage (A V scale) 1. Set knob Sk2 to position “ “. 2. Set kno
17、b Sk1 to position “1200 V“ (highest voltage range, to prevent overloading) and connect the measuring leads to the sockets + and . 3. Now connect the measuring leads to the voltage source to be measured and set knob Sk1 to that measuring position that is most suitable to provide an accurate reading o
18、f the voltage in question. If the direct voltage in question has a maximum value of .60 mV, knob Sk2 may be put in position “60 mV “. (Sk1 may have any position; the meter protection is cut off!) B. Direct current (A V scale) 1. Set knob Sk2 to position “ “. 2. Set knob Sk1 to position “3 A“ (highes